YSI MultiLab IDS 4320 Conductivity and Temperature Sensor
IDS 4320 digital conductivity/temperature probe for the YSI MultiLab. Features a stainless steel body, 2-electrodes, and a 1.5 meter cable. The IDS 4320 probe is designed for ultra-pure water applications and cannot be calibrated by the user, although the probe is factory calibrated.
- Features
- Specifications
The IDS sensors automatically store their unique serial number and calibration data. In addition, they also digitally process the measurement signal. The IDS sensors can be moved from instrument to instrument and maintain their calibration data and transmit that information to the new instrument.
Specification | |
---|---|
Cell Constant |
0.100 cm-1 ± 2 % |
Equipment used with | MultiLab Instruments |
Measurement Range |
Conductivity: 0.01 µS/cm to 200 µS/cm | Resistivity: 5 kOhm cm to 100 MOhm cm | Temperature: 0 to 100ºC |
Minimum Immersion Depth | 30 mm |
Parameters Measured | Conductivity, Temperature |
Storage Temperature | 0 to 50 °C |
Warranty |
2 years |